The
Bidentate C18 shows greater selectivity between the critical peak pair at
higher temperature whereas the UDC shows the opposite trend. Since shape
selectivity effects are known to be more pronounced when the UDC stationary
phase moiety is more rigid, it makes sense that the lower temperatures would
produce greater separation between the analytes. The Bidentate C18 stationary
phase on the other hand does not have this property and therefore the effect is
not the same. In fact, when using the
Bidentate C18 at 15 °C where the UDC had shown the best separation, bexarotene
appeared as a shoulder peak on the tazarotene peak.
Peak
efficiency initially increased with increasing temperature until it reached a
maximum and then decreased (see Figure 3). The drop-off was more significant on
the UDC column, suggesting that rigidity of the UDC moiety may play a more
important role in the retention than the Bidentate C18. The maximum values were
obtained in the range of 35–40 °C.
Knowing
how temperature affects the separation can lead to more informed choices during
method development. Resolution depends on both efficiency and selectivity, but
their optimum values are not always the same; the selectivity on the UDC was
best at 15 °C but the efficiency was highest at 35 °C. The analyst should
consider the relative importance of each effect that the temperature has on the
chromatographic data when choosing an optimum column temperature.
Table 1. Method conditions for the separations.
Parameter
|
Details
|
Solvent A
|
DI water + 0.1% formic acid
|
Solvent B
|
Acetonitrile + 0.1% formic acid
|
Flow Rate
|
1.0 mL/min
|
Gradient
|
0–1 min hold at 30%B, 1–24min to 100%B, 24–25min to 30%B
|
Detection
|
UV 254nm
|
Injection Volume
|
10µL
|
Figure 1. Retention as a function of temperature on the UDC
column.
Figure 2. Retention as a function of temperature on the
Bidentate C18 column.
Figure 3. Efficiency of tazarotene as a function of
temperature on both columns.
No comments:
Post a Comment