…Rule out other possibilities first! Problems you can encounter
with peak shape, retention, and so on can often be traced to a component of the
HPLC system. For example, consider the following real-world situation which we encountered
in our laboratory.
We were
running an HPLC method for ethylbenzene (0.1mg/mL concentration) with a Cogent
UDA™ column and observed some problems with the chromatograms (see Figure A). We
noticed that our analyte peak was retained longer and longer after each
injection and that the peak became broader. Was this changing chromatographic
behavior due to column damage? We know the columns are very stable so we
investigated other more likely possibilities first.
We were
using a 0.5mL/min isocratic mobile phase from two solvent reservoirs (70%
solvent A, 30% solvent B). Thinking there may be a problem with one of the solvent
lines, we prepared a premixed mobile phase instead and only used one mobile
phase reservoir (solvent A). Under these conditions, we obtained the results in
shown in Figure B. Notice how the retention time precision is now very robust
and the efficiency is high. This demonstrates that the column performance is
fine and the problem was with the solvent B line. Perhaps the issue was at the
mixing stage, in which the solvent proportioning valve was not mixing the
correct amount of solvent from the B line.
Also
apparent from Figure A was a second, smaller peak that sometimes would show up
in the chromatograms and sometimes would not. As before, we investigated system
issues as the possible cause. It turned out that it was a problem with the
injector. We were using an injection volume of 1 µL, and when we increased it
to 5 µL, the peak was consistently present in every run (Figure B).
So
before you blame the column, be sure to inspect everything else thoroughly:
• Check your mobile
phase. Are the A and B lines connected to the correct inlet ports? Has the
flow rate been calibrated for each line? Is the solvent mixer proportioning
valve functioning accurately? When was the mobile phase prepared and does it
need to be replaced?
• Check your method. Are
you using an appropriate mobile phase/gradient for your compounds? Use an ANP
gradient for polar compounds and a reversed phase one for less polar analytes.
Are you operating at a proper detection wavelength for your compounds’ UV
absorption properties? Before running a sequence, look through your method
settings and make sure someone else who used the instrument before you didn't accidentally change anything.
• Check your sample
prep. Are the analyte concentrations suitable? Ensure that overload of the
column or the detector will not occur at these concentrations and dilute if
necessary. Have your samples been properly filtered to remove particulate
matter from entering the system and creating blockages? Have you performed
percent recovery studies to demonstrate that the analytes are completely
present in the final sample for analysis?
• Check your injector.
Is the needle going down far enough to reach the liquid sample? Is the syringe
aspirating the correct volume of sample?
This is of course not a comprehensive troubleshooting list
but it can give you some idea of the type of things that can malfunction. An
HPLC analysis can be complex and involves many factors so do not start with
blaming the column.